HOME取扱製品エレクトリカルコンポーネントパルスジェネレータQuantum Composers

9730シリーズ パルス電流ジェネレータ

  • 拡大

    9730シリーズ パルス電流ジェネレータ

  • 正面

  • 背面

9730 シリーズは、エアバッグや起爆剤テスト用途に開発されました。

シンプルな操作性、短時間での設定変更、プログラム機能などの特長を持つ、高い精度及び再現性の必要な用途に最適なシステムです。

エアバッグ開発(小型起爆)や点火装置開発において、信頼性、安全性、再現性を提供します。

  • Up to 4 individual outputs with fully individual programming and control.
  • Up to 6A output per channel.
  • Front and rear external trigger inputs.
  • Current monitor – The current output from the unit is monitored and a voltage representation the waveform is presented at a front panel BNC connector.
  • Voltage monitor- The voltage across the load being driven is monitored and a voltage representation is present at a front panel BNC connector.
    *Instantaneous power and total energy may be derived with the aid of both the current and voltage monitors.
  • SAFETY - Keyed front panel fire enable, mechanical and software interlocks.
  • Complete channel and system setup stored in memory. Provides 12 memory storage slots.
  • Meets Bruceton and Neyer requirements.
  • Analog signals depicting current and voltage.

標準仕様
  • 200 ns Timing Resolution
  • 2 or 4 Channel Outputs
  • 6 amps per channel
  • Easy Programming Interface
  • Integrated Bridgewire Resistance Measurement
  • Safety Interlocks and Keyed Front Panel Interface
  • 12 Memory Storage Slots
  • TTL Sync output to T0
  • Full Customer Support
  • 2 Year Warranty
Congurations
9732 – 2 Output Channels
9734 – 4 Output Channels

input module - front and rear trigger

INTERNAL RATE GENERATOR
rate 0.01 Hz to 40 KHz
resolution 100 ns
accuracy 20 ns
jitter 10 ns RMS
burst mode 2 to 250 pulses
output modes single pulse, burst
control modes internal rate generator, external trigger

EXTERNAL TRIGGER INPUTS
 
function generate individual pulses (single shot or burst)
front or rear selectable between front or rear panel inputs
external trigger rate Min = 5 X Longest Pulse Width
Max = 200 Seconds
insertion delay 300 ns
jitter 10 ns
impedance 1K Ω
Slope rising or falling
trigger filter filters out unwanted “glitch” or “runt” pulses
range 0.02 us to 1 ms
resolution 0.02 us
trigger level 2-15 V
level threshold 100 mV

RISE TIMES
rising edge - inductive 4 us (1 ohm, 50’ cable, 6 A) varies with load
rising edge - resistive 550 ns (1 ohm, 2” cable, 6 A)varies with load

SLEW RATES
rising edge - inductive
(varies with load)
2 A/us @ 6 A
1 A/us @ 3 A
0.30 A/us @ 1 A
(1 ohm and 50’ cable)
falling edge - inductive
(varies with load)
-1.60 A /us @ 6 A
-0.90 A/us @ 3 A
-0.30 A/us @ 1 A
(1 ohm and 50’ cable)
rising edge - resistive
(varies with load)
6.3 A/us @ 6 A
9.5 A/us @ 3 A
2.2 A/us @ 1 A
(1 ohm & 2” cable)
falling edge - resistive
(varies with load)
-5.2 A/us @ 6 A
-4.0 A/us @ 3 A
-1.3 A/us @ 1 A
(1 ohm & 2” cable)

OUTPUTS
amplitude 0.10 – 6 A
resolution 2 mA
accuracy +/- 0.5% - 2%
compliance voltage level 19 - 22 Volts
pulse width range 5 us – 100 s *
error +/- 0.1 %
resolution 100 ns
delay range 0-30 s
error +/- 0.1 %
* Maximum pulse width is limited by current amplitude. 1A can go up to 100 s and 6 A is limited to 300 ms.

MONITOR OUTPUTS
voltage monitor (isolated) 0.2 V/V
current monitor (isolated) 0.5 V/A
error <4%
bandwidth Min = 100 kHz
Max = 1000kHz

RESISTANCE MEASUREMENT
range 0.1 to 150 Ohms
resolution .01 Ohms
error .5 to 15 Ohms < 4%
error 16 to 150 Ohms <10%
measurement current 100 mA max.

GENERAL
synch outputs, front and rear T0, channels A through D
standard communications usb (serial bridge), RS232 (115200, 57600,
38400, 19200, 9600, 4800 baud)
size 9” 2 U x 10” 2U size rack mount
electrical 100-240 V, 50-60 Hz

SAFETY
remote interlock shorting interlock
arming key switch removable keyswitch
internal error checking checks control circuit for errors

OPTIONS
E- ethernet
  • Integrated 4-wire resistance measurement for each channel with pre and post testing features. Ability to set the mean and upper/lower limits to determine what construes a resistance fault.
  • Current and voltage monitor outputs.
  • Front and rear sync outputs.
  • Safety features include remote interlock, removable keyed enable switch and internal error checking.
  • RS232, USB, and optional Ethernet computer programming interfaces.